Assessment of the immunity to radiated disturbances of an FPAA-based front-end device for ECG signal monitoring

Dalila Salhi, Nicola Pasquino
Abstract:
The paper is focused on the assessment of the functional immunity of an FPAA circuit configured to operate as a front-end element to be used in ECG signal conditioning, and on the investigation of the possible effects of electromagnetic disturbance on the parameters of the ECG signal, within a range of frequency from 250 MHz to 3 GHz. The research is carried out with a methodology which relies on typical ECG signal and frequency domain analysis.
Keywords:
electromagnetic compatibility, electromagnetic interference, FPAA, immunity measurements, ECG, biomedical application
Download:
IMEKO-TC4-2013-147.pdf
DOI:
-
Event details
IMEKO TC:
TC4
Event name:
TC4 Symposium 2013
Title:
19th IMEKO TC4 Symposium Measurements of Electrical Quantities (together with 17th TC4 IWADC Workshop on ADC and DAC Modelling and Testing)
Place:
Barcelona, SPAIN
Time:
18 July 2013 - 19 July 2013