A fully automated calibration system for pressure balance

T. Kobata

Abstract

A new fully automated system for calibrating a pressure balance is described. The system consists of two automated pressure balances, a pressure controller, two air-operated constant volume valves, a precise pressure transducer, a device for measuring environmental condition, a computer and the program developed. In the system, one pressure balance is used as a standard and the other is calibrated and/or evaluated. To determine the equilibrium state between two pressure balances accurately, the cross-float measurement using the comparator method is performed. In this paper, the details of the system are described and the results obtained using the system are shown.

Keywords
pressure balance, calibration, cross-float, automated system