Skip to main content

High pressure comparison among seven European national laboratories

M. Rantanen, S. Semenoja and K. Riski

Abstract

An inter-comparison in the high oil pressure range was arranged in 2005 – 2007. The participating laboratories were CMI / Czech Republic, Metas / Switzerland, MIKES / Finland, NMi-VSL / The Netherlands, PTB / Germany, SMD / Belgium and SP / Sweden. The transfer standard was a Desgranges & Huot piston-cylinder unit with a nominal effective area of 1,96 mm², mounted in a D&H 5316 balance body. The participants determined the effective area at ten nominal pressures from 50 MPa to 500 MPa in steps of 50 MPa. All the results were in a good agreement with the reference values, calculated as weighted means of the results from the four primary level laboratories. The results fully support the uncertainties claimed in the CMC tables of the participants.

Download
IMEKO-TC16-2007-045u.pdf
DOI
-
IMEKO TC
TC16 - Pressure and Vacuum Measurement

Event details

Event
Cultivating Metrological Knowledge
Technical Committee
TC16
Place
Merida, MEXICO
Time
27 November 2007 - 1 December 2007
Website
http://www.cenam.mx/fyp/imeko2.html

Back to the proceedings