Skip to main content

The long-term behaviour of force transducers as a criteria for the selection of new transfer standards

Aimo Pusa

Abstract

The laboratory of MIKES-Lahti Precision Oy has made over almost twenty years comparisons with the PTB, utilizing more than ten different force transducers. These comparisons have been made regularly every year to supervise the traceability of force. The results of these comparisons are showing different kinds of behaviours. For the traceability the most important characteristic is the stability of the sensitivity. The change of the sensitivity directly influences the measurement uncertainty and as a result affects the final uncertainty of the intercomparison. The second characteristic is the zero reading, which can indicate the change of the measurement behaviour of the transducer.

Keywords
force transducer, long term stability, force intercomparison
Download
IMEKO-TC3-2007-057u.pdf
DOI
-
IMEKO TC
TC3 - Measurement of Force, Mass, Torque, and Gravity

Event details

Event
Cultivating Metrological Knowledge
Linked User
Dirk Röske
Technical Committee
TC3
Place
Merida, MEXICO
Time
27 November 2007 - 1 December 2007
Website
http://www.cenam.mx/fyp/imeko2.html

Back to the proceedings