Expansion of the calibration range and improvement of the uncertainty in the 1 kN·m deadweight torque standard machine
Abstract
Expansion of the calibration range and improvement of the uncertainty of realized torque in the 1 kN·m deadweight torque standard machine were investigated. The lower limit of the calibration range was lowered from 5 N·m to 0.5 N·m by developing new small linkage weight series. In addition, the sensitivity limit at the fulcrum was reevaluated. The moment-arm length was also re-measured after changing the thin metal bands at the ends of the moment-arm. As a result of these evaluations, relative expanded uncertainties (k = 2) of 7.3 × 10<sup>-5</sup> and 2.9 × 10<sup>-5</sup> could be obtained in the range from 0.5 N·m to 20 N·m and from 5 N·m to 1 kN·m, respectively.
Event details
- Event
- Cultivating Metrological Knowledge
- Linked User
- Dirk Röske
- Technical Committee
- TC3
- Place
- Merida, MEXICO
- Time
- 27 November 2007 - 1 December 2007
- Website
- http://www.cenam.mx/fyp/imeko2.html