The UME “M One” mass comparator system was installed at the UME to be used research activities in describing an optimum transfer standard as buoyancy artifacts and comparing buoyancy artifacts between in air and vacuum environments and also 1 kg mass standards. The system were designed and manufactured by Mettler Toledo. The reproducibility of the mass comparator is 0,4 µg under vacuum.