Skip to main content

NEW EXPERIENCES WITH A FORCE MEASURING FACILITY FROM THE RANGE FROM 1 mN TO 5 N

S. Niehe

Abstract

In this contribution new experiences with a force measuring facility [1] for the range from 1 mN to 5 N consisting of a piezoelectric positioning unit for force generation and a precision compensation balance for force measuring are presented. The enhancement of the facility by means of a rotational table allows to rotate the force transducer under test around its axis and to measure the sensitivity of the transducer at different positions. By averaging the measured sensitivities of a complete rotation cycle the rotation effect can be taken into account and the deviation from the calibration of the investigated force transducer in a deadweight force standard machine is then of the order of 10<sup>-4</sup> in relative terms. This improves the deviation of earlier results [1] without consideration of rotation effects about a factor of 8.

Download
IMEKO-TC3-2005-024u.pdf
DOI
-
IMEKO TC
TC3 - Measurement of Force, Mass, Torque, and Gravity

Event details

Event
Force, Mass and Torque Measurements
Linked User
Dirk Röske
Technical Committee
TC3
Email
imeko@nlab.org
Place
Cairo, EGYPT
Time
19 February 2005 - 23 February 2005
Website
http://www.nlab.org/imeko/

Back to the proceedings