Skip to main content

COMPARISON MEASUREMENTS OF MIKES-RAUTE 20 kN·m TORQUE REFERENCE DEVICE WITH THE PTB

Aimo Pusa, Dirk Röske, Michael Sachs

Abstract

This article deals with the vertical 20 kN·m torque reference calibration device newly built by MIKES- RAUTE. It widens the possibilities of the MIKES-RAUTE Mass and Force Laboratory in the field of torque calibrations from 2 kN·m up to 20 kN·m. A brief overview about the construction and setup of the device is included in this report. Also some information about the applied calibration process is given. The main part of this article describes comparison measurements with the PTB, Germany, from summer 2004 and presents their results. They show that the device will attain an uncertainty level of 5·10<sup>-4</sup> or better over the entire measurement range.

Download
IMEKO-TC3-2005-033u.pdf
DOI
-
IMEKO TC
TC3 - Measurement of Force, Mass, Torque, and Gravity

Event details

Event
Force, Mass and Torque Measurements
Linked User
Dirk Röske
Technical Committee
TC3
Email
imeko@nlab.org
Place
Cairo, EGYPT
Time
19 February 2005 - 23 February 2005
Website
http://www.nlab.org/imeko/

Back to the proceedings