THE USE OF A µCMM IN THE CALIBRATION OF HARDNESS INDENTERS

O Kruger, L Mostert

Abstract

In order to perform traceable hardness measurements using Rockwell Hardness Indenters, the indenter must conform to the requirement of ISO 6508-3 standard. The standard defines specifications for the geometric parameters for this type of indenter. Conformance to this standard is of particular importance to all calibration laboratories which operate in accordance with ISO/IEC 17025. At present there is no facility in South Africa with the capability of calibrating indenters for conformance to ISO 6508-3. This situation is being addressed through a series of studies currently being conducted at the National Metrology Institute of South Africa. The following parameters are calibrated using the µCMM. ISO 6508-3 specifies a radius of 200 µm with an uncertainty of ± 5 µm for Rockwell Hardness Indenter tips. Also specified in the ISO standard is the cone perpendicularity between the axis of the diamond cone to the axis of the indenter holder, with a specification of ± 0,03°. The last part of the ISO standard which was investigated states: “The surface of the cone and the spherical tip shall blend in a truly tangential manner.” All these parameters were measured and associated uncertainties calculated to prove conformance with ISO 6508-3 for the calibration of hardness indenters.

Keywords
Rockwell hardness, ISO 6508-3, µCMM
Download
IMEKO-TC5-2007-017.pdf
DOI
-
IMEKO TC
TC5 - Hardness Measurement

Event details

Event
Symposium on Recent Advancements in the Theory and Practice of Hardness Measurement
Technical Committee
TC5
Place
Tsukuba, JAPAN
Time
19 November 2007 - 21 November 2007
Website
http://www.hardmeko2007.org

Back to the proceedings