Skip to main content

CALIBRATION OF THE DEPTH-MEASURING DEVICE OF ROCKWELL HARDNESS TESTING MACHINES

F. Loeffler (F. Löffler), A. Sawla, P. Strobel

Abstract

In addition to the indirect verification using reference blocks and a reference indenter, it is possible to realise the calibration by direct verification of the test force, the indenter, the depth-measuring device and the testing cycle. Some conditions of the calibration process, e.g. the verification range, and the necessary accuracy of the depth-measuring device are described in standard ISO 6508-2. With reference to these parameters, a new depthmeasuring calibration device was developed. The study describes the function of this device and the calibration process. The advantages of this new process are that the measurement takes place in the axis of force application and that the calibration is independent of the Rockwell hardness scales. During calibration, the force stresses the calibration device but not the sensitive measuring sensor. Results of a real depth-measuring device verification are also presented.tion process. The advantages of this new process are that the measurement takes place in the axis of force application and that the calibration is independent of the Rockwell hardness scales. During calibration, the force stresses the calibration device but not the sensitive measuring sensor. Results of a real depth-measuring device verification are also presented.

Download
IMEKO-TC5-2002-008.pdf
DOI
-
IMEKO TC
TC5 - Hardness Measurement

Event details

Event
Joint International Conference on Force, Mass, Torque, Hardness and Civil Engineering Metrology in the age of globalization
Technical Committee
TC5
Place
Celle, GERMANY
Time
24 September 2002 - 26 September 2002

Back to the proceedings