A NEW WAY TO MINIMIZE UNCERTAINTY IN CALIBRATION PROCESS OF FORCE TESTING MACHINES

Boris Katz

Abstract

The results of the calibration of load cells (force transducers) according to ISO 376-1999 are described in table of values where force <i>F</i> (kN) is a function of electric signal <i>R</i> (mV/V) and by analytical approximation of the calibration curve.
The analytical approximation is usually expressed by polynom of 2nd or 3rd degree. The measurement of force in the calibration of testing machines by means of digital unit device DMP-40 (HBM) suppose application of linear interpolation between two adjacent values of signal <i>R</i>.
By means of mathematical calculations it is theoretically proven in which cases the linear interpolation is undesirable and leads to additional interpolation error, and increasing of the value of uncertainty measurement in accordance.
A few studies and experiments of the issue confirm the theoretical conclusions: in some cases minimizing interpolation error and uncertainty measurement in measuring force F as a function of reading the signal <i>R</i>, can be achieved by approximation polynom instead of linear interpolation between two adjacent values. For this purpose QCC Hazorea has developed program called “MABA-2000,” which can solve this problem.

Keywords
materials testing machines, measurements uncertainty, interpolation error, calibration curve
Download
IMEKO-TC3-2002-012.pdf
DOI
-
IMEKO TC
TC3 - Measurement of Force, Mass, Torque, and Gravity

Event details

Event
Joint International Conference on Force, Mass, Torque, Hardness and Civil Engineering Metrology in the age of globalization
Technical Committee
TC3
Place
Celle, GERMANY
Time
24 September 2002 - 26 September 2002

Back to the proceedings