Skip to main content

UNCERTAINTY IN TORQUE CALIBRATION USING VERTICAL TORQUE AXIS ARRANGEMENT AND SYMMETRICAL TWO FORCE MEASUREMENT

Dieter Kenzler, Jörg Oblasser, Andreas Subaric-Leitis, Christian Ullner

Abstract

Calibration machines up to a torque range of 20 kN&middot;m need extensive design if the force is performed by dead weights. A machine based on two force reference standards and a vertical torque axis may serve to avoid these specific problems. The precision force transducers work at a lever arm of precision length. The test piece is mounted in vertical arrangement between a hydraulically operating rotary-actuator and the torque measuring system. Such a calibration machine allows additionally performing easily continuous calibration. Regarding an overall relative uncertainty not greater than 2&middot;10<sup>-4</sup> there are some significant problems. The estimation of the uncertainty has been proved by comparative tests with the national torque standard by the PTB according to the guidelines EA-10/14 and EA-2/03.

Download
IMEKO-TC3-2002-018.pdf
DOI
-
IMEKO TC
TC3 - Measurement of Force, Mass, Torque, and Gravity

Event details

Event
Joint International Conference on Force, Mass, Torque, Hardness and Civil Engineering Metrology in the age of globalization
Technical Committee
TC3
Place
Celle, GERMANY
Time
24 September 2002 - 26 September 2002