In the present paper, we discuss the integration, characterization and operation of the recently established an automated 10 volt Josephson Series Array Voltage Standard (JSAVS) at National Physical Laboratory, India (NPLI). Furthermore, we discuss the precision calibration of the Zener reference standard alongwith the uncertainty budget. The combined standard uncertainty of the Zener reference standard calibration at 10V for a Single Point Measurement (SPM) [1] is ± 25 nV at k=1.