Skip to main content

CALIBRATION OF 100 MΩ HAMON RESISTOR USING CURRENT-SENSING WHEATSTONE BRIDGE

Ivan Leniček, Roman Malarić, Alan Šala

Abstract

Primary electromagnetic laboratory in Zagreb maintains standards of four major electromagnetic units: volt, farad, ohm and second. Reference standards are periodically calibrated in PTB (Germany) and several methods has been realized for internal laboratory traceability check. RC-comparison method uses Hamon divider constructed of 100 equally balanced 1-megaohm resistors that connected in parallel or serially exhibit Hamon ratio of 100 M&#937; : 10 k&#937; with relative uncertainty of 10<sup>-8</sup>. Hamon resistance of 100 M&#937; is directly compared with impedance of 100 pF capacitance reference at frequency of 15,873 Hz. Measurement of low-frequency voltage ratio is based on two digital voltmeters HP 3458A with improved algorithm for data processing. Accuracy of reference standards comparison 10 k&#937; &#8594;100 pF depends on consistency of Hamon divider that can be disturbed by aging of used metal-film components. In this paper a recently realized method for accurate measurement of 100 M&#937; Hamon resistor by means of comparison with calibrated and stable 10 M&#937; resistance standard at higher voltage levels has been presented.

Download
IMEKO-TC4-2004-020.pdf
DOI
-
IMEKO TC
TC4 - Measurement of Electrical Quantities

Event details

Event
TC4 Symposium 2004
Technical Committee
TC4
Email
imeko@mail.ntua.gr
Place
Athens, GREECE
Download
imeko_tc4_2004_call_for_papers2.pdf
Time
29 September 2004 - 1 October 2004
Website
http://www.medialab.ntua.gr/IMEKO

Back to the proceedings