Skip to main content

A CENTRONICS BASED TRANSDUCER INDEPENDENT INTERFACE (TII) FULLY COMPLIANT WITH 1451.2 STD.

Helena Ramos, M. Pereira, V. Viegas, O. Postolache, P. Girão

Abstract

The paper reports the tests conducted on a dedicated Transducer Independent Interface (TII) to check its compliance with standard IEEE 1451.2 specifications. The interface was built using a synchronous data transfer based on the Serial Peripheral Interface (SPI) protocol integrated in a PIC microcontroller used as the Smart Transducer Interface Module (STIM) and the Centronics port of the PC, used as the Network Capable Application Processor (NCAP’s TII port). The description of each line and the connections are also presented.<br />The working prototype of the STIM controlled by a NCAP module was implemented to connect several transducers without communication capabilities used in a water quality monitoring network. The NCAP, also described in the paper, is emulated in a PC and it was fully developed using LabVIEW 6.1.

Download
IMEKO-TC4-2004-097.pdf
DOI
-
IMEKO TC
TC4 - Measurement of Electrical Quantities

Event details

Event
TC4 Symposium 2004
Technical Committee
TC4
Email
imeko@mail.ntua.gr
Place
Athens, GREECE
Download
imeko_tc4_2004_call_for_papers2.pdf
Time
29 September 2004 - 1 October 2004
Website
http://www.medialab.ntua.gr/IMEKO