Skip to main content

ELECTROMAGNETIC IMMUNITY OF A PORTABLE DATA ACQUISITION SYSTEM

Salvatore Nuccio, Ciro Spataro, Giovanni Tinè

Abstract

In the paper the performance of a portable data acquisition system under radiated electromagnetic disturbances is investigated, in order to find out if and how its peculiar characteristics change with respect to the ones measured in absence of electromagnetic field. The analysis is carried out by means of experimental tests, subjecting the system under test to the threats considered by the IEC-61326 standard and following the test procedures prescribed in the IEC-61000-4-3 standard. The results show that the electromagnetic influence can lead to a worsening of the tested data acquisition system features.

Keywords
Electromagnetic Compatibility, Analog-to-Digital Conversion, Data Acquisition System
Download
PWC-2003-TC4-098.pdf
DOI
-
IMEKO TC
TC4 - Measurement of Electrical Quantities

Event details

Event
XVII IMEKO World Congress
Place
Dubrovnik, CROATIA
Time
22 June 2003 - 28 June 2003
Website
http://www.hmd.hr/imeko/

Back to the proceedings