Skip to main content

RADIATION ISODOSE SURFACE DISTORTION AS A SOURCE OF DOSE OR EXPOSURE RATE MEASUREMENT UNCERTAINTY: EXAMPLE IN BRACHYTHERAPY SEEDS

José Manzoli, Jorge Pirolla, Eduardo Moura, Carlos Zeituni, João Moura, Maria Elisa Chuery Martins Rostelatto

Abstract

Radiation sealed sources have an emission that could be studied by theirs isodose surfaces. The intersection of a plan with such a surfaces is described by the dose rate contour curves. It is very common to assume that these surfaces have rotational symmetry around the longitudinal axis, and so, the contour curves have mirror symmetries. Here it is shown, by a computational simulation, that if the radioactive material inside de sealed vessel is not homogeneous, the uncertainty in dose rate measurements could achieve high values. This uncertainty value is exemplified in a iodine-125 seed source used for brachytherapy purposes.

Keywords
radiation, sealed source, isodose
Download
IMEKO-WC-2009-TC13-069.pdf
DOI
-
IMEKO TC
TC13 - Measurements in Biology and Medicine

Event details

Event
XIX IMEKO World Congress
Place
Lisbon, PORTUGAL
Time
6 September 2009 - 11 September 2009
Website
http://www.imeko2009.it.pt/

Back to the proceedings