Skip to main content

NOVEL AND LOW-COST TEMPERATURE COMPENSATION TECHNIQUE FOR PIEZORESISTIVE PRESSURE SENSORS

Ferran Reverter, Goran Horak, Vedran Bilas, Manel Gasulla

Abstract

This paper proposes a low-cost technique to compensate for the temperature dependence of piezoresistive pressure sensors configured in a Wheatstone bridge. The sensor is treated as a resistive circuit and three equivalent resistances are measured by setting appropriately the bridge terminals. One of these equivalent resistances depends on temperature but not on pressure and, hence, it can be used to compensate for the temperature dependence of the output parameter. In such a way, neither data of a previous temperature calibration nor additional components are required to compensate for the temperature dependence. The proposed technique is applied to a commercial pressure sensor, which is measured by means of a direct sensor-to-microcontroller interface circuit. Experimental results show that temperature effects on the pressure measurement decrease more than ten times when the proposed technique is applied.

Keywords
pressure sensor, piezoresistive sensor, temperature compensation, sensor electronic interface, microcontroller
Download
IMEKO-WC-2009-TC16-074.pdf
DOI
-
IMEKO TC
TC16 - Pressure and Vacuum Measurement

Event details

Event
XIX IMEKO World Congress
Place
Lisbon, PORTUGAL
Time
6 September 2009 - 11 September 2009
Website
http://www.imeko2009.it.pt/

Back to the proceedings