Skip to main content

TRACEABILITY OF 633 nm LASER CALIBRATION AT NIMT

Monludee Ranusawud, Ketsaya Vacharanukul, Anusorn Tonmueanwai

Abstract

The National Institute of Metrology (Thailand), NIMT, maintains the standard of length in accordance with the definition of the meter through a 633 nm iodine stabilized He-Ne laser. Its accuracy is transferred to stabilized and non-stabilized lasers regarding to beat-frequency method and direct measurement using a wavelength meter, respectively. The traceability in wavelength measurement at the NIMT is discussed here. The iodine stabilized He-Ne laser is self-calibrated using the master and slave beat frequency system. NIMT participated in APMP.L-K11 intercomparison and the results are presented. The standard uncertainty is 2.5 × 10<sup>-11</sup>.

Keywords
iodine stabilized He-Ne laser, beat frequency method, traceability
Download
IMEKO-WC-2009-TC8-158.pdf
DOI
-
IMEKO TC
TC8 - Traceability in Metrology

Event details

Event
XIX IMEKO World Congress
Place
Lisbon, PORTUGAL
Time
6 September 2009 - 11 September 2009
Website
http://www.imeko2009.it.pt/