Skip to main content

VOLUME RATIO DETERMINATION IN STATIC EXPANSION SYSTEMS BY MEANS OF TWO PRESSURE BALANCES

David Herranz, Salustiano Ruiz, Mª. Nieves Medina

Abstract

Static expansion systems are used to generate known pressures in the high and medium vacuum range. In these systems the volume ratios are the crucial parameters and have to be determined very accurately. For this task different methods can be used. We have used the gas accumulation method with a dead weight pressure balance, a force-balanced piston gauge (FPG) and a differential pressure transducer (CDG 133.32 Pa). The final uncertainty of the expansion ratios obtained with this method is similar to other institutes uncertainties.

Keywords
vacuum, expansion
Download
IMEKO-WC-2009-TC16-280.pdf
DOI
-
IMEKO TC
TC16 - Pressure and Vacuum Measurement

Event details

Event
XIX IMEKO World Congress
Place
Lisbon, PORTUGAL
Time
6 September 2009 - 11 September 2009
Website
http://www.imeko2009.it.pt/