Skip to main content

ACCURATE PICOSCALE FORCES FOR INSITU CALIBRATION OF AFM

Koo-Hyun Chung, Gordon Shaw, Jon R. Pratt

Abstract

The force sensitivity of an atomic force microscope is calibrated directly using an in situ realization of primary electrostatic forces ranging from 320 pN to 3.3 nN with accuracy of a few percent. The absolute accuracy of a common atomic force microscope (AFM) force calibration scheme, known as the thermal noise method is evaluated via comparison to the electrostatic calibration. It is demonstrated that the thermal noise method can be applied with great success to yield force measurements with relative standard uncertainties below 5 % after application of a geometric correction factor.

Keywords
electrostatics, atomic force microscope, piconewton
Download
IMEKO-WC-2009-TC3-285.pdf
DOI
-
IMEKO TC
TC3 - Measurement of Force, Mass, Torque, and Gravity

Event details

Event
XIX IMEKO World Congress
Place
Lisbon, PORTUGAL
Time
6 September 2009 - 11 September 2009
Website
http://www.imeko2009.it.pt/

Back to the proceedings