Skip to main content

DEVELOPMENT OF A FLOW-THROUGH CELL FOR ACCURATE MEASUREMENTS OF LOW ELECTROLYTIC CONDUCTIVITY

Chiara Boveri, Francesca Durbiano, Danilo Serazio

Abstract

The present paper focuses on the development of a flow-through cell and a closed circuit that permit to carry out measurements of low electrolytic conductivity of aqueous solutions under flowing condition. The traceability path has been set as follows: samples with conductivity between 200 mS cm<sup>-1</sup> and 50 mS cm<sup>-1</sup> have been employed for the calibration of the geometric constant of the new cell, by comparison with the primary cell of Istituto Nazionale di Ricerca Metrologica (I.N.Ri.M.). Then the flow-through cell has been used to measure values with decreasing conductivities down to 1 mS cm<sup>-1</sup>. The measurement system capabilities have been evaluated to be limited by contamination effects.

Keywords
pure water, low conductivity measurement, flow-through cell
Download
IMEKO-WC-2009-TC24-450.pdf
DOI
-
IMEKO TC
TC24 - Chemical Measurements

Event details

Event
XIX IMEKO World Congress
Place
Lisbon, PORTUGAL
Time
6 September 2009 - 11 September 2009
Website
http://www.imeko2009.it.pt/

Back to the proceedings