Abstract
A high-precision multi-range unbalanced transformer bridge designed for inductance standard calibration is described. Basic theory of the bridge is discussed which enables direct measurement of differences of both impedance components (Δ<i>L</i> and Δ<i>R</i>). The instrument was designed to compare inductors from 100 mH to10 H in the frequency range from 100 Hz to10 kHz. The uncertainty of comparison in the impedance range from 1 W to 10 kW is assessed to be smaller than 10 ppm.
Event details
- Event
- XIX IMEKO World Congress
- Place
- Lisbon, PORTUGAL
- Time
- 6 September 2009 - 11 September 2009
- Website
- http://www.imeko2009.it.pt/