LEEB HARDNESS-CALIBRATION-MACHINE
Abstract
For the calibration of hardness reference blocks according to the German standard DIN 50156 -hardness test according Leeb -MPA NRW has realized a new concept of hardness reference standard machine. A laser interferometer is used for the continuous measurement of distance and time. In a post process the speed of the indenter will be evaluated from the measurement data. The conception of the machine, measurement capabilities and the effects of variation of test parameters are reported in the following paper.
Event details
- Event
- HARDMEKO 2010
- Technical Committee
- TC5
- Place
- Pattaya, THAILAND
- Time
- 21 November 2010 - 25 November 2010
- Website