PREPARATION PROCEDURE FOR LOADING FLUX QUANTUM WEIGHTS ON PICONEWTON FORCE REALIZATIION SYSTEM
Abstract
We present a preparation procedure for loading flux quantum weights on piconewton force standard system, being developed at KRISS, including the micropositioning of a cantilever device with a superconducting micro-ring. To locate the device at zero-magnetic-field position inside a high-field-gradient superconducting magnet, the behaviour of the mechanical resonance frequency of the device in magnetic fields were studied in a dynamic mode and used to probe the local magnetic field value at the device position. A vertical offset of a few hundred micrometers caused a shift of ~ 0.7 Hz to the resonance frequency of 1144.4 Hz, and was nulled by either additional uniform magnetic field or a <i>z</i>-axis nanopositioner. Nulling was also confirmed by a non-singular temperature dependence of the resonance frequency.
Event details
- Event
- XX IMEKO World Congress
- Place
- Busan, REPUBLIC of KOREA
- Time
- 9 September 2012 - 12 September 2012
- Website
- http://imeko2012.kriss.re.kr