Skip to main content

INTER-LABORATORY COMPARISON OF REFERENCE TORQUE WRENCH CALIBRATION BETWEEN NMIJ AND PTB IN THE RANGE OF FROM 200 N·m TO 2 kN·m

Koji Ogushi, Atsuhiro Nishino, Koji Maeda, Kazunaga Ueda, Andreas Brüge, Dirk Röske

Abstract

Inter-laboratory comparison of reference torque wrench calibration was conducted between NMIJ and PTB using a torque transducer in the form of a torque wrench, "TTS/3kNm", as a transfer device. Although the rated capacity of the transducer is 3 kN·m, the calibration range of from 200 N·m to 2 kN·m was selected. Each loading time schedule used in daily calibration was also adopted in this comparison. Calibration results obtained by both laboratories coincided with the range of uncertainties.

Keywords
torque, torque wrench, reference torque wrench, comparison, square drive
Download
IMEKO-WC-2012-TC3-O32.pdf
DOI
-
IMEKO TC
TC3 - Measurement of Force, Mass, Torque, and Gravity

Event details

Event
XX IMEKO World Congress
Place
Busan, REPUBLIC of KOREA
Time
9 September 2012 - 12 September 2012
Website
http://imeko2012.kriss.re.kr

Back to the proceedings