Skip to main content

AN ECONOMIC WAY TO HAVE SMALL TEMPERATURE VARIATION CONTROLLED IN MASS MEASUREMENT

Feng-yu Yang, Sheau-shi Pan, Sheng-Jui Chen

Abstract

In APMF2011, we introduced a simple way by building a two-layer insulation chamber with two mass comparators inside to have small temperature variation in our laboratory. In this paper, we will introduce a economic method by adding a simple active control unit on the aluminum chamber to control the temperature inside the chamber at (20 ± 0.1) °C nearly, while the room temperature of the air-conditioning laboratory is active controlled at (21.5 ± 1.5) °C. Using this method, the transient phenomenon of temperature was observed.

Keywords
temperature variation / fluctuation, passive / active control, two-layer insulation chamber
Download
IMEKO-WC-2012-TC3-P22.pdf
DOI
-
IMEKO TC
TC3 - Measurement of Force, Mass, Torque, and Gravity

Event details

Event
XX IMEKO World Congress
Place
Busan, REPUBLIC of KOREA
Time
9 September 2012 - 12 September 2012
Website
http://imeko2012.kriss.re.kr