DC MAGNETIC FLUX DENSITY STANDARDS IN THE RANGE OF EARTH MAGNETIC FIELD

Po Gyu Park, Young Gyun Kim, Wan-Seop Kim, Mun-Seog Kim, V. Ya. Shifrin

Abstract

We have carried out precision magnetic measurements utilizing a magnetic-resonance technique, aiming for improving the magnetic-field standard below 100 mT. The standard for measuring dc magnetic flux density is intended to cover the range from 10 mT to 100 mT, with an uncertainty (<i>k</i> = 2) ranging from 4 mT/T to 10 mT/T.

Keywords
magnetic flux density, magnetic resonance
Download
IMEKO-WC-2012-TC4-O12.pdf
DOI
-
IMEKO TC
TC4 - Measurement of Electrical Quantities