UNCERTAINTY EVALUATION FOR THE CALIBRATION OF DIRECT ACTING INDICATING ANALOGUE ELECTRICAL MEASURING INSTRUMENTS

T. Goto

Abstract

In the case of calibration of direct acting indicating analogue electrical measuring instruments (hereinafter called the "analogue instruments"), uncertainty budget largely depends on dispersion of readings for scales through a needle. This paper describes experimental results for three levels of voltage (10 V, 6 V, and 3 V) of analogue instruments. In the experiments, two way layout nested design was introduced for each voltage level. From the results of present experiments, it was revealed that expanded uncertainties obtained were about 1/3 for the specification that obtained from the past experience.

Keywords
direct acting indicating analogue measuring instruments, nested design, ANOVA
Download
IMEKO-WC-2012-TC4-O28.pdf
DOI
-
IMEKO TC
TC4 - Measurement of Electrical Quantities

Event details

Event
XX IMEKO World Congress
Place
Busan, REPUBLIC of KOREA
Time
9 September 2012 - 12 September 2012
Website
http://imeko2012.kriss.re.kr

Back to the proceedings