In this paper is presented the method for comparison of high value resistance standards by using the modified Wheatstone bridge and picoammeter as current detector, instead of low-impedance current-sensing transimpedance amplifier. Basically, the resistance ratio <i>r</i> can be measured on nominal levels 1:1, 10:1 or 100:1, ensuring the traceability chain from the reference standards of 10 M&Omega; and 1 G&Omega; for the range of interest 10 M&Omega; to 100 G&Omega;.

Keywords
calibration, high resistance standards, zero-current measurement, modified Wheatstone bridge
Download
IMEKO-WC-2012-TC4-O8.pdf
DOI
-
IMEKO TC
TC4 - Measurement of Electrical Quantities

Event details