A PRECISION DC LOW-CURRENT SOURCE FOR CALIBRATION OF ELECTROMETERS IN THE RANGE OF 100 pA TO 1 fA
Abstract
A precision DC low-current source based on charging a capacitor with a linearly varying voltage ramp was built to calibrate low-current meters in the range of 100 pA to 1 fA. For a calibration of Keithley 6430, the relative expanded measurement uncertainty (<i>k</i> = 2) was evaluated to be 16 µA/A ~ 62 mA/A within the current range. In the calibration process, it was observed that the offset current of the capacitors lead to a crucial contribution to the measurement uncertainty.
Event details
- Event
- XX IMEKO World Congress
- Place
- Busan, REPUBLIC of KOREA
- Time
- 9 September 2012 - 12 September 2012
- Website