Skip to main content

A PRECISION DC LOW-CURRENT SOURCE FOR CALIBRATION OF ELECTROMETERS IN THE RANGE OF 100 pA TO 1 fA

Wan-Seop Kim, Mun-Seog Kim, Dan bee Kim, Young Gyun Kim, Po Gyu Park

Abstract

A precision DC low-current source based on charging a capacitor with a linearly varying voltage ramp was built to calibrate low-current meters in the range of 100 pA to 1 fA. For a calibration of Keithley 6430, the relative expanded measurement uncertainty (<i>k</i> = 2) was evaluated to be 16 µA/A ~ 62 mA/A within the current range. In the calibration process, it was observed that the offset current of the capacitors lead to a crucial contribution to the measurement uncertainty.

Keywords
DC low-current, voltage ramp, capacitor, picoammeter
Download
IMEKO-WC-2012-TC4-P20.pdf
DOI
-
IMEKO TC
TC4 - Measurement of Electrical Quantities

Event details

Event
XX IMEKO World Congress
Place
Busan, REPUBLIC of KOREA
Time
9 September 2012 - 12 September 2012
Website
http://imeko2012.kriss.re.kr

Back to the proceedings