DEVELOPMENT OF ATOMIC MAGNETOMETER SYSTEM APPROACHING TO THE SPIN EXCHANGE RELAXATION FREE REGIME
Abstract
We describe a potassium atomic magnetometer based on the spin exchange relaxation free regime. The optical rotation signal was optimized under the condition of temperature of 200 °C and width was about 0.15 mG when the pump laser power was 55 mW. Signal-to-noise measurements yield a sensitivity of 120 fT/Hz<sup>1/2</sup> at 10 Hz.
Event details
- Event
- XX IMEKO World Congress
- Place
- Busan, REPUBLIC of KOREA
- Time
- 9 September 2012 - 12 September 2012
- Website