Skip to main content

CONTACT-FREE FLOW RATE MEASUREMENT IN WEAKLY CONDUCTING ELECTROLYTES USING LORENTZ FORCE VELOCIMETRY

A. Wegfrass, C. Diethold, M. Werner, C. Resagk, F. Hilbrunner, B. Halbedel, A. Thess

Abstract

The feasibility of Lorentz Force Velocimetry for flow measurement of poorly conducting electrolytes is demonstrated using experimental results on salt water flow exposed to a permanent magnet system. The results provide a linear relationship between the measured Lorentz force and flow velocity, consistent with the theoretical scaling laws. Further measurements are currently being performed on a new experimental setup. The new experiments are aimed at investigating the effect of flow dynamics on the measured force.

Keywords
Lorentz force, weakly conducting electrolytes, flow rate measurement
Download
IMEKO-WC-2012-TC9-O4.pdf
DOI
-
IMEKO TC
TC9 - Flow Measurement

Event details

Event
XX IMEKO World Congress
Place
Busan, REPUBLIC of KOREA
Time
9 September 2012 - 12 September 2012
Website
http://imeko2012.kriss.re.kr

Back to the proceedings