Skip to main content

STUDY ON THERMAL DRIFT OF NOBLE METAL THERMOCOUPLES

Sung-Sin So, Hyun-Min Park, Yeun-Joon Jung

Abstract

We set up the temperature - running period CM (Characteristic Modelling) of the noble metal thermocouples (S/B type). Using CM, we got the both thermocouples max error reduced, type S is 2.1 °C and type B is 1.8 °C which were 5.8 °C and 4.3 °C before using CM and the coefficients of determination in CM are 82.1 % and 72.5 %. It shows 85 % difference between CM and error less than 1.0 °C, especially in type B. So it is necessary to use CM to measure temperature accurately.

Keywords
thermal drift, characteristic modelling, type S thermocouple, type B thermocouple
Download
IMEKO-WC-2012-TC12-O14.pdf
DOI
-
IMEKO TC
TC12 - Temperature and Thermal Measurements

Event details

Event
XX IMEKO World Congress
Place
Busan, REPUBLIC of KOREA
Time
9 September 2012 - 12 September 2012
Website
http://imeko2012.kriss.re.kr

Back to the proceedings