Skip to main content

FETAL BIOMETRY IN KOREAN FETUS

Sung Eun Hur, Sung Ki Lee, B. Dongjoo Shin, Hyun Kyoon Lim, Se Jin Park

Abstract

The purpose of the study was to create a fetal biometry standard from data on ultrasound-dated in Korean people. A total of 142 singleton gestations with ultrasonographic fetal biometry from 14 to 38 weeks made up the study population. BPD, HC, AC and FL were measured every second weeks. To assess sonographic accuracy, we measured wooden sticks in 0°, 10° and 70° in artificial box. BPD, HC, AC and FL were measured. Mean values of each percentiles (5, 25, 50, 75, 95) are displayed. The resulting values were plotted graphically. Inter-degree differences were not significant and 0.2 ± 0.1, 0.3 ± 0.2, 0.2 ± 0.1, respectively (0° VS 10°, 0° VS 70° and 10° VS 70°) We validated the accuracy and reliability of sonographic measurements according to fetal positions.

Keywords
biometry, ultrasonography, fetus
Download
IMEKO-WC-2012-TC13-O10.pdf
DOI
-
IMEKO TC
TC13 - Measurements in Biology and Medicine

Event details

Event
XX IMEKO World Congress
Place
Busan, REPUBLIC of KOREA
Time
9 September 2012 - 12 September 2012
Website
http://imeko2012.kriss.re.kr

Back to the proceedings