HIGH-RESOLUTION AUTOCOLLIMATOR CALIBRATION SYSYTEM

H. Fujimoto, Watanabe

Abstract

This paper describes high-resolution autocollimator calibration system in NMIJ. A goniometer system with nano-radian resolution provides precision rotation angles of a mirror attached on the rotation axis and rotation angles read by an autocollimator are compared to the set angles. The rotation angle is detected by a high-precision angle interferometer that observes motions of corner reflectors attached to the axis. The interferometer is calibrated to output absolute angles by a 24-sided optical polygon that is attached to the rotation axis. The performance of the system is verified by comparing the calibration curves of same autocollimator taken by this system and those taken by the national standard of NMIJ.

Keywords
autocollimator, angle interferometer, X-ray diffraction, optical polygon
Download
IMEKO-WC-2012-TC14-O7.pdf
DOI
-
IMEKO TC
TC14 - Measurement of Geometrical Quantities

Event details

Event
XX IMEKO World Congress
Place
Busan, REPUBLIC of KOREA
Time
9 September 2012 - 12 September 2012
Website
http://imeko2012.kriss.re.kr