Skip to main content

PRIMARY VACUUM STANDARD CMI FOR UHV PRESSURE RANGE 1 × 10<sup>-6</sup> TO 1 × 10<sup>-10</sup> Pa

T. Gronych, L. Peksa, M. Jeráb, Z. Krajícek, D. Pražák, F. Stanek, M. Vicar, J. Tesar

Abstract

A primary vacuum standard for UHV pressure range is under development in Czech Metrology Institute (CMI). In order to cover the UHV range down to the limit 1 × 10<sup>-10</sup> Pa a method of calibration in non-equilibrium gas (so called dynamic extension method) was employed. The apparatus is divided into three parts: the dynamic extension stage, the orifice flow standard and the flowmeter. The dynamic extension stage is designed as an XHV apparatus manufactured from a low outgassing rate material – beryllium copper alloy. A special 4 K bakeable refrigerator cryopump is used as the main pump.

Keywords
primary vacuum standard, vacuum gauge calibration in UHV range
Download
IMEKO-WC-2012-TC16-P4.pdf
DOI
-
IMEKO TC
TC16 - Pressure and Vacuum Measurement

Event details

Event
XX IMEKO World Congress
Place
Busan, REPUBLIC of KOREA
Time
9 September 2012 - 12 September 2012
Website
http://imeko2012.kriss.re.kr

Back to the proceedings