The efficiency of a <sup>226</sup>Ra aqueous solution source in the radium-radon secular equilibrium against the gamma reference ionization chamber in KRISS has been obtained both by calculating from the photon energy-dependent efficiency curve of the ionization chamber and by measuring with an <sup>226</sup>Ra standard source. The two results of calculation and measurement have showed such a good agreement that the gamma reference ionization chamber may be used to certificate the radioactivity of <sup>226</sup>Ra aqueous solutions in the secular equilibrium and its traceability could be linked to the multi-radionuclide calibration of the ionization chamber.