Skip to main content

ELECTRONIC SKIN REPLACES VOLUNTEER EXPERIMENTS

Norbert Nessler, Werner Reischer, Marcel Salchner

Abstract

Neutral electrodes (grounding plates used in electrosurgical operations) were tested with thermo cameras in volunteers experiments. These experiments are difficult to perform and expensive. A new test device swaroTEST with a measuring board "electronic skin" simulates the relevant electrical features of human skin to replace volunteer experiments. This device consists of a three dimensional resistor network representing the electric features of the skin and muscle tissue, and a temperature sensing array (one transistor for each cm²) to measure the resultant temperature increase after a standardized current load (700 mA hf current during 60 s, proposed in the relevant AAMI HF-18 standard). The comparison of test results with thermo camera images of volunteer experiments showed a sufficient coincidence, which proves the applicability of this test device to replace volunteer experiments.

Keywords
electronic skin, electrosurgery, neutral electrode
Download
PWC-2003-TC10-006.pdf
DOI
-
IMEKO TC
TC10 - Measurement for Diagnostics, Optimization and Control

Event details

Event
XVII IMEKO World Congress
Place
Dubrovnik, CROATIA
Time
22 June 2003 - 28 June 2003
Website
http://www.hmd.hr/imeko/