Skip to main content

UNCERTAINTY ANALYSIS OF CELL COUNTING BY METABOLIC ASSAYS

C.Divieto, G.Sassi, M.P.Sassi

Abstract

Cell counting is an important procedure in cell biology and its uncertainty evaluation is fundamental to improve cell-based measurement comparability which is currently variable. In this study, for the first time, the uncertainty of a metabolic assay for cell counting based on fluorescence intensity measurement has been evaluated. Results, in 2D cell culture, revealed that three are the main components: repeatability and reproducibility of the measurement system and a contribution due to the sample loading step procedure. This study represents the basis of a 3D cell culture in which metabolic methodologies are the most suitable method to monitor cell growth.

Keywords
2D and 3D cell culture, scaffold, cell proliferation rate, CellTiter Blue, resazurin, uncertainty analysis
Download
IMEKO-WC-2012-SS2-O11.pdf
DOI
-

Event details

Event
XX IMEKO World Congress
Place
Busan, REPUBLIC of KOREA
Time
9 September 2012 - 12 September 2012
Website
http://imeko2012.kriss.re.kr