ERROR EFFECTS IN MICROLASER SENSORS
Abstract
Microlaser sensors use the intracavity photoelastic effect in Nd:YAGlaser crystals to detect force and force-related quantities like torque, pressure and inertial acceleration. Error effects are caused by local and temporal temperature changes as well as by nonlinear influences of the laser crystal, its support and the load system. Measurement resolution is limited by seismic noise and nonreproducibility. We have achieved a low relative temperature coefficient of 10<sup>-6</sup>/°C and a relative reproducibility of 10<sup>-5</sup> related to a maximum input load of 100 N. These results are advantageous properties compared with some competing force sensors.
Event details
- Event
- XVI IMEKO World Congress
- Place
- Vienna, AUSTRIA
- Time
- 25 September 2000 - 28 September 2000