VECTOR DC MAGNETOMETERS CALIBRATION ERROR
Abstract
The calibration system for vector DC magnetometers of Nurmijarvi Geophysical Observatory (Finland) is described. It is based on a three axes reference coil system and Observatory reference magnetometer. It is shown experimentally that the system allows to determine calibrated magnetometer parameters with an error about 0.02%. One important error source -influence of noncollinearity of reference and calibrated magnetometer axes - is analyzed and value of this error is estimated. Also additional possibilities of the Observatory infrastructure are described: determination of temperature errors of magnetometer’s sensor and electronics and long-term temporal drift study of up to 9 magnetometers simultaneously.
Event details
- Event
- XVI IMEKO World Congress
- Place
- Vienna, AUSTRIA
- Time
- 25 September 2000 - 28 September 2000