Skip to main content

VECTOR DC MAGNETOMETERS CALIBRATION ERROR

V. Korepanov, F. Dudkin, K. Pajunpää

Abstract

The calibration system for vector DC magnetometers of Nurmijarvi Geophysical Observatory (Finland) is described. It is based on a three axes reference coil system and Observatory reference magnetometer. It is shown experimentally that the system allows to determine calibrated magnetometer parameters with an error about 0.02%. One important error source -influence of noncollinearity of reference and calibrated magnetometer axes - is analyzed and value of this error is estimated. Also additional possibilities of the Observatory infrastructure are described: determination of temperature errors of magnetometer’s sensor and electronics and long-term temporal drift study of up to 9 magnetometers simultaneously.

Keywords
magnetometer, calibration, axes collinearity, error
Download
IMEKO-WC-2000-TC4-P137.pdf
DOI
-
IMEKO TC
TC4 - Measurement of Electrical Quantities

Event details

Event
XVI IMEKO World Congress
Place
Vienna, AUSTRIA
Time
25 September 2000 - 28 September 2000

Back to the proceedings