Skip to main content

INVESTIGATION ON THE REALIZATION OF AN AUTOMATED AND GUARDED HAMON 10×10 GΩ NETWORK

Flavio Galliana, Pier Paolo Capra, E. Gasparotto

Abstract

Recently at National Institute of Metrological Research (INRIM) a Hamon actively guarded 10 × 100 MΩ was developed to improve the traceability levels at 1 GΩ level. Utilizing and revisiting this project, a Hamon 10 × 10 GΩ network is now under realization to extend the capabilities of the Hamon method up to 100 GΩ. Moreover, by means of this Hamon network, it will be possible to extend the capabilities of the measurement method for the calibration of picoammeters already developed at INRIM.

Download
IMEKO-TC4-2010-036.pdf
DOI
-
IMEKO TC
TC4 - Measurement of Electrical Quantities

Event details

Event
Instrumentation for the Information and Communication Technology Era
Technical Committee
TC4
Place
Kosice, SLOVAKIA
Time
8 September 2010 - 10 September 2010
Website
http://www.imeko.tuke.sk

Back to the proceedings