ON THE METHODOLOGIES FOR THE CALIBRATION OF STATIC ELECTRICITY METERS IN THE PRESENCE OF HARMONIC DISTORTION

Antonio Cataliotti, Valentina Cosentino, Alessandro Lipari, Salvatore Nuccio

Abstract

This paper is focused on the problems related to the calibration of static electricity meters with nonsinusoidal test waveforms. This is an up-to-date issue since the harmonic distortion levels are ever increasing, especially in low-voltage power systems, and in such conditions the performances of the modern static electricity meters can be negatively affected, depending on the implemented measurement algorithm. Thus it is important to test the meters with nonsinusoidal waveforms. The paper analyzes the problems related to both the choice of the test signals to be used for calibration and the implemented metric, with a particular focus on the reactive energy meters. The analysis is supported by several experimental results.