Skip to main content

Development of coating thickness gauge for magnetic coatings

Jae-Hwa Hong, Jung Hun Park

Abstract

A gauge to measure the coating thickness of magnetic coatings on non-magnetic base material has been developed by directly measuring the magnetic flux density between the coating and the permanant magnet. Magnetic induction method was found to be inadequate for measuring magnetic coating thickness which is greater than 0.4 mm. Magentic flux density was measured by Hall effect sensor and calibration curve was constructed by using the data acquired from 10 standard reference samples. The developed gauge can be used to measure non-magnetic coatings over magnetic substrate as well. The relative standard error was found to be 2.1% for nickle coatings on copper substrate.

Download
IMEKO-TC10-2013-014.pdf
DOI
-
IMEKO TC
TC10 - Measurement for Diagnostics, Optimization and Control

Event details

Event
TC10 Workshop 2013
Technical Committee
TC10
Place
Florence, ITALY
Time
6 June 2013 - 7 June 2013
Website
http://www.imekotc10-florence.org/

Back to the proceedings