Interlaboratory Comparison of Microwave Power Measurement Using Type N Coaxial Line
Abstract
An interlaboratory comparison of microwave power measurement (project UME-EM-10-02) was carried out among eight laboratories in Turkey. In this comparison, it was requested to make microwave power measurements at +10 dBm, 0 dBm and -15 dBm power levels with 0.03 GHz, 1 GHz, 7 GHz, 12 GHz and 18 GHz frequencies with their own measurement systems and methods. The measurement results and the expanded uncertainties were presented by the participant laboratories. Also, the normalized score (<i>E</i><sub>n</sub>) of each participant laboratory were calculated and presented to evaluate the performance of the laboratories. Most of the results in measuring power on a type N coaxial line show good agreement. According to the calculated <i>E</i><sub>n</sub> values, corrective actions were proposed to some of the participant laboratories.
Event details
- Event
- TC4 Symposium 2013
- Technical Committee
- TC4
- joaquin.del.rio@upc.edu
- Place
- Barcelona, SPAIN
- Time
- 18 July 2013 - 19 July 2013
- Website
- http://www.imeko2013.es/