Skip to main content

Water Fraction Measurement Using a RF Resonant Cavity Sensor

Heron Eduardo Ávila, Daniel Juan Pagano, Fernando Ragel Sousa

Abstract

We have tested the working principle of water fraction measurement using a Radio Frequency (RF) resonant cavity sensor. This technology allows us to determine the water fraction by the frequency shift of the first resonant peak in a non-intrusive way. The sensor operates with low power signals (1 mW) in the range of 100 through 400 MHz. Static measurements were done using a Scalar Network Analyzer (SNA) for water/oil and water/air static mixtures with fresh water and sea water. The measurement’s uncertainty and the characteristic curves of the sensor’s behavior were computed.

Keywords
water fraction measurement, resonant cavity sensor, radio frequency
Download
IMEKO-TC4-2013-069.pdf
DOI
-
IMEKO TC
TC4 - Measurement of Electrical Quantities

Event details

Event
TC4 Symposium 2013
Technical Committee
TC4
Email
joaquin.del.rio@upc.edu
Place
Barcelona, SPAIN
Time
18 July 2013 - 19 July 2013
Website
http://www.imeko2013.es/

Back to the proceedings