The harmonic content generated by the use of non-linear loads can cause an electrical device to operate improperly. For this reason, reliable measurements of this parameter are fundamental to guarantee the results needed by the electronics industry and medical devices. The aim of this work is to describe the development of a measurement system of harmonic signals with traceability to the international system of units (DC voltage), and its uncertainty levels. The system consist a software developed in Visual Basic language based on a voltmeter with a General Purpose Interface Bus (GPIB) able to digitalize a signal with sampling rates up to 100,000 samples per second. A Discrete Fourier Transform algorithm is used to evaluate the harmonic content of the signal. The work consists of a tool for the calibration services provided by the laboratory LABELO-PUCRS, and its results indicate the best capability (uncertainty) to harmonic measurements.

Keywords
Harmonics, Calibration, Uncertainty, DFT
Download
IMEKO-TC4-2013-083.pdf
DOI
-
IMEKO TC
TC4 - Measurement of Electrical Quantities

Event details

Event
TC4 Symposium 2013
Technical Committee
TC4
Email
joaquin.del.rio@upc.edu
Place
Barcelona, SPAIN
Time
18 July 2013 - 19 July 2013
Website
http://www.imeko2013.es/