Skip to main content

AN INTERNATIONAL COMPARISON OF AC JOSEPHSON VOLTAGE STANDARDS BETWEEN NRC, CANADA AND VSL, THE NETHERLANDS

Piotr S. Filipski, Helko van den Brom, Ernest Houtzager

Abstract

The paper presents results of an international comparison of two quantum AC Josephson Voltage Standards based on pulse- driven Josephson arrays. The two systems differ in several hardware and software characteristics as well as in the level of automation, features which can influence the accuracy of transferring the quantum -standard voltage value to a calibrated instrument. The comparison was conducted at 100mV, 20mV and 12mV, at frequencies between 2.5kHz and 100kHz. An electronically-aided thermal transfer standard was used as a travelling standard. At the most accurate voltage and frequency point, 100mV at 2.5kHz, both laboratories agreed to better than 1 part in 106.

Keywords
Josephson arrays, signal synthesis, standards, ac voltage standard
Download
IMEKO-TC4-2011-018.pdf
DOI
-
IMEKO TC
TC4 - Measurement of Electrical Quantities

Event details

Event
TC4 Symposium 2011
Technical Committee
TC4
Email
eventos@metrologia.org.br
Place
Natal, BRAZIL
Time
27 September 2011 - 30 September 2011
Website
http://www.metrologia.org.br/metrologia2011/

Back to the proceedings