Skip to main content

ECG SIMULATOR FOR TESTING AND SERVICING CARDIAC MONITORS AND ELECTROCARDIOGRAPHS

Samuel E. de Lucena

Abstract

This paperwork presents the design and test of a microcontroller-based ECG (electrocardiogram) simulator. The ECG simulator is capable of generating nine ECG signals and a calibrated square wave (1 Hz, 1 mV at Lead II). The synthesized signals are: normal sinus rhythm (60 beats per minute – BPM, and 90 BPM), sinus bradycardia (30 BPM), sinus tachycardia (120, 180 and 240 BPM), sinus rhythm with tall T wave (6EC0 BPM), ventricular tachycardia (120 BPM), and a rhythm for asynchronous ventricular pacemaker (60 BPM). The simulated ECG signals can have their amplitudes continuously varied from zero to 2 mV at lead II, through a panel potentiometer, whose knob position does not affect the amplitude of the square wave calibrated signal. The technique used to synthesize the ECG signals, a modified direct digital synthesis (DDS), appeared superior in permitting the generation of signals of very good quality and yet using a relatively small amount of memory. This latter feature enabled the use of a simple 8-bit microcontroller together with an external precision D/A converter and a 2-pole lowpass active filter. The designed ECG simulator can be a handsome instrument to daily ECG machine test, maintenance and demonstration by hospital personnel and medical equipment maintenance groups and shops.

Keywords
ECG simulators, electrocardiographs, cardiac monitor, maintenance, medical instrumentation
Download
IMEKO-TC4-2011-024.pdf
DOI
-
IMEKO TC
TC4 - Measurement of Electrical Quantities

Event details

Event
TC4 Symposium 2011
Technical Committee
TC4
Email
eventos@metrologia.org.br
Place
Natal, BRAZIL
Time
27 September 2011 - 30 September 2011
Website
http://www.metrologia.org.br/metrologia2011/

Back to the proceedings